| Specifications |
4320 |
4520 |
| Input |
| Voltage |
115/230V AC ± 10%, User Selection |
115/230V AC Auto Range, ± 15% variation |
| Frequency |
50/60 Hz ±5% |
50/60 Hz ±5% |
| Fuse |
6.3A 250V Slow blow |
15A 250V Slow blow |
| Dielectric Withstand Test mode |
| Output Rating |
5 kV @ 20 mAAC |
5 kV @ 100 mAAC |
| |
6 kV @ 5 mADC |
6 kV @ 10 mADC |
| Voltage Setting |
| - Range |
0.00 - 5.00 kVAC |
0.00 - 5.00 kVAC |
| |
0.00 - 6.00 kVDC |
0.00 - 6.00 kVDC |
| - Resolution |
0.01 kV |
0.01 kV |
| - Accuracy |
± (2% of setting + 5 volts) |
± (2% of setting + 5 volts) |
| Voltage Display |
| - Range |
0.00 - 5.00 kVAC |
0.00 - 5.00 kVAC |
| |
0.00 - 6.00 kVDC |
0.00 - 6.00 kVDC |
| - Resolution |
0.01 kV |
0.01 kV |
| - Accuracy |
± (2% of reading + 10 volts) |
± (2% of reading + 10 volts) |
| Current Display |
| - Range |
0.00 - 20.00 mAAC |
0.00 – 99.99 mAAC |
| |
0.00 - 5.00 mADC |
0.00 – 10.00 mADC |
| - Resolution |
0.01 mA |
0.01 mA |
| - Accuracy |
± (2% of reading + 0.02 mA) |
± (2 % of reading + 0.06 mA) |
| Hi - Lo Limit AC |
| - Range |
0.00 - 20.00 mA |
0.00 – 99.99 mAAC |
| - Resolution |
0.01 mA |
0.01 mA |
| - Accuracy |
± (2% of setting + 0.02 mA) |
(2 % of reading + 0.06 mA) |
| Hi - Lo Limit DC |
| - Range |
0.00 - 5.00 mA |
0.00 – 10.00 mADC |
| - Resolution |
0.01 mA |
0.01 mA |
| - Accuracy |
± (2% of setting + 0.02 mA) |
(2 % of reading + 0.06 mA) |
| |
|
|
| Failure Detector |
Audible and Visual |
Audible and Visual |
| DC Output Ripple |
< 5% at 6 kVDC @ 5 mA, resistive load |
≤ 5% Ripple RMS at 6 KVDC @ 10 mA, resistive load |
| Discharge Time |
≤ 200 ms |
≤ 200 ms |
| Max. Capacitive Load in DC Mode |
| - 1.00 uF |
< 1 kV |
< 1 kV |
| - 0.75 uF |
< 2 kV |
< 2 kV |
| - 0.50 uF |
< 3 kV |
< 3 kV |
| - 0.08 uF |
< 4 kV |
< 4 kV |
| - 0.04 uF |
< 5 kV |
< 5 kV |
| - 0.01 uF |
< 6 kV |
< 6 kV |
| |
|
|
| AC Wave Form |
Sine Wave, Crest Factor = 1.3 – 1.5 |
Sine Wave, Crest Factor = 1.3 – 1.5 |
| AC Output Frequency (Range) |
60 or 50 Hz, user selection |
60 or 50 Hz, user selection |
| Output Regulation |
± (1 % of setting + 5 volts) from no load to full load |
± (1 % of setting + 5 volts) from no load to full load |
| Dwell Timer |
| - Range |
0, 0.2 - 999.9 sec (0 = continuous) |
0, 0.2 - 999.9 sec (0 = continuous) |
| - Resolution |
0.1 sec increments |
0.1 sec increments |
| - Accuracry |
± (0.1% + 0.05 sec) |
± (0.1% + 0.05 sec) |
| Ramp Timer |
|
|
| - Range |
0.1 – 999.9 sec |
0.1 – 999.9 sec |
| - Resolution |
0.1 sec increments |
0.1 sec increments |
| - Accuracry |
± (0.1% + 0.05 sec) |
± (0.1% + 0.05 sec) |
| |
|
|
| Insulation Resistance Test Mode |
| Output Voltage |
|
|
| - Range |
100 – 1000 Volts DC |
100 – 1000 Volts DC |
| - Resolution |
1 volt |
1 volt |
| - Accuracry |
± (2% of reading + 5 volts) |
± (2% of reading + 5 volts) |
| Voltage Display |
|
|
| - Range |
0 – 1000 V |
0 – 1000 V |
| - Resolution |
1 volt |
1 volt |
| - Accuracry |
± (2% of reading + 2 counts) |
± (2% of reading + 2 counts) |
| Resistance Display |
|
|
| - Range |
1 - 1000 MΩ (4 digit, auto ranging) |
1 - 1000 MΩ (4 digit, auto ranging) |
| - Resolution 500VDC, 0.01MΩ |
1.00–40.00 |
1.00–40.00 |
| - Resolution 500VDC, 0.1MΩ |
35.0–999.9 |
35.0–999.9 |
| - Resolution 1000VDC, 0.01MΩ |
1.00–80.00 |
1.00–80.00 |
| - Resolution 1000VDC, 0.1MΩ |
75.0–999.9 |
75.0–999.9 |
| HI-Limit Range |
0, 1 – 1000 MΩ (0 = off) |
0, 1 – 1000 MΩ (0 = off) |
| LO-Limit Range |
1 – 1000 MΩ |
1 – 1000 MΩ |
| |
0, 0.5 - 999.9 sec (0 = constant) |
0, 0.5 - 999.9 sec (0 = constant) |
| Delay Timer |
|
|
| - Resolution |
0.1 sec |
0.1 sec |
| - Accuracy |
± (0.1% + 0.05 sec) |
± (0.1% + 0.05 sec) |
| |
|
|
| Ground Bond Test Mode |
| Output Voltage Range |
6 VAC fixed |
6 VAC fixed |
| Output Frequency Range |
60 or 50 Hz, user selectable |
60 or 50 Hz, user selectable |
| Output Current |
| - Range |
0.1 Amp |
0.1 Amp |
| - Resolution |
± (2 % of setting + 0.02 A) |
± (2 % of setting + 0.02 A) |
| - Accuracry |
0.0 – 30.0 Amps |
0.0 – 30.0 Amps |
| Current Display |
| - Range |
0.0 – 30.0 Amps |
0.0 – 30.0 Amps |
| - Resolution |
0.1 Amp |
0.1 Amp |
| - Accuracry |
± (3 % of reading + 0.1 A) |
± (3 % of reading + 0.1 A) |
| HI-Limit LO-Limit |
| - Range |
0 – 510 mΩ for 3.0 – 10.0 Amps |
0 – 510 mΩ for 3.0 – 10.0 Amps |
| |
0 – 200 mΩ for 10.1 – 25.0 Amps |
0 – 200 mΩ for 10.1 – 25.0 Amps |
| |
0 – 150 mΩ for 25.1 – 30.0 Amps |
0 – 150 mΩ for 25.1 – 30.0 Amps |
| - Resolution |
1 mΩ |
1 mΩ |
| - Accuracy |
± (2 % of setting + 2 mΩ) |
± (2 % of setting + 2 mΩ) |
| Dwell Timer |
| - Range |
0, 0.5 - 999.9 sec (0 = constant) |
0, 0.5 - 999.9 sec (0 = constant) |
| - Resolution |
0.1 sec |
0.1 sec |
| - Accuracry |
± (0.1% + 0.05 sec) |
± (0.1% + 0.05 sec) |
| Milliohm Offset |
| - Max. Offset Capability: |
100 mΩ |
100 mΩ |
| - Resolution |
1 mΩ |
1 mΩ |
| - Accuracy |
± (2 % of setting + 2 mΩ) |
± (2 % of setting + 2 mΩ) |
| |
|
|
| General Specifications |
| PLC Remote Control |
Inputs – test, reset, interlock & recall memory 1–6 |
| Outputs – pass, fail, test-in-process |
| Memory |
Allows storage of up to 6 different test programs and 6 steps per memory and a single step mode |
| Interface |
Optional RS-232 interface |
| Security |
Lockout capability to avoid unauthorized access to test set-up programs |
| Calibration |
Software & adjustments made through front panel |
| Mechanical |
Bench or rack mount with tilt up front feet |
| Dimensions (W x H x D) |
11” x 3.5” x 17” (280 x 89 x 430 mm) |
16.9” x 5.2” x 15.7” (430 x 133 x 400 mm) |
| Weight |
33 lbs. (15 kg) |
54 lbs. (24.5 Kg) |
| |
|
| Supplied Accessories |
|
| 102-050-913 |
High Voltage Probe (6 ft.) |
| 102-055-913 |
Input Voltage Clip (6 ft.) |
| 125-013-001 |
Input Power Cable (6 ft.) (4320) |
| 99-10164-01 |
Input Power Cable (6 ft.) (4520) |
| 99-10008-01 |
High Current Return Lead |
| 99-10009-01 |
High Current Output Lead |
| 99-10040-01 |
Interlock |
| 99-10106-01 |
Fuse (4320) |
| 99-10656-01 |
Fuse (4520) |